12.5
µM
625
mg/mL
12,500
nM
625
mg
12.5
nmol
12.5
µM
625
mg/mL
12,500
nM
625
mg
12.5
nmol
The Protein Concentration Calculator determines protein concentration from UV absorbance measurements using the Beer-Lambert law. UV spectrophotometry at 280 nm is the standard method for quantifying purified proteins, offering speed, simplicity, and non-destructive measurement. This calculator converts raw absorbance readings into molar (µM, nM) and mass (mg/mL) concentrations.
Accurate protein concentration measurement is critical for enzyme kinetics, binding assays, crystallography trials, drug formulation, and any quantitative biochemical experiment. The calculator accommodates custom path lengths for microvolume spectrophotometers, dilution factors, and computes total protein mass in the sample volume for downstream applications.
The Beer-Lambert law relates absorbance to concentration:
$$A = \varepsilon \times c \times l$$
Rearranging for molar concentration:
$$c = \frac{A}{\varepsilon \times l}$$
where $$A$$ is the measured absorbance (unitless), $$\varepsilon$$ is the molar extinction coefficient (M⁻¹cm⁻¹), $$c$$ is concentration (M), and $$l$$ is the optical path length (cm). The mass concentration is then:
$$c_{mg/mL} = c_M \times MW \times 10^{-3}$$
If the sample was diluted before measurement, the actual concentration is:
$$c_{actual} = c_{measured} \times \text{Dilution Factor}$$
The Beer-Lambert law is valid in the absorbance range of approximately 0.1 to 1.0. Below 0.1, signal-to-noise is poor; above 1.0, deviation from linearity occurs due to detector limitations and molecular interactions.
The calculated concentration assumes the protein is pure and the extinction coefficient is accurate. An A₂₈₀ reading between 0.1 and 1.0 falls within the linear range of the Beer-Lambert law, providing the most reliable results. If your absorbance exceeds 1.5, dilute the sample and apply the dilution factor.
The A260/A280 ratio indicates sample purity: pure protein gives ~0.57, nucleic acid contamination raises this above 0.7. For impure samples, use the Warburg-Christian correction or alternative quantification methods. Note that buffer components, detergents (especially those with aromatic rings like Triton X-100), and reducing agents can absorb at 280 nm and interfere with measurements.
Inputs
Results
An IgG antibody (ε₂₈₀ = 210,000, MW = 150 kDa) with A₂₈₀ = 0.75 at 10x dilution gives a stock concentration of ~35.7 µM or ~5.36 mg/mL.
Inputs
Results
A 35 kDa recombinant protein at A₂₈₀ = 0.35 (undiluted) gives 12.5 µM or 0.44 mg/mL. With 2 mL total, this is 0.875 mg (25 nmol) of protein.
The Beer-Lambert law is most accurate between A = 0.1 and A = 1.0. Below 0.1, the signal-to-noise ratio is poor, leading to high relative error. Above 1.0, most spectrophotometers deviate from linearity due to stray light and detector saturation. Ideally, aim for A₂₈₀ between 0.2 and 0.8 for the best precision (typically less than 2% error).
Three approaches: (1) Calculate from sequence using the Pace method (Trp × 5500 + Tyr × 1490 + Cystine × 125), available from ExPASy ProtParam, (2) Look up in literature or protein databases, (3) Determine experimentally by measuring A₂₈₀ of a sample with known concentration (from amino acid analysis or quantitative amino acid hydrolysis).
For mixtures, A₂₈₀ gives total protein concentration only if you use an average extinction coefficient. This is unreliable because different proteins have very different ε₂₈₀ values. For mixtures, colorimetric assays like BCA or Bradford provide more consistent results because they measure total peptide bonds or amino acids regardless of aromatic content.
NanoDrop spectrophotometers use 0.05-1 mm path lengths and require only 1-2 µL of sample. They internally correct to 1-cm equivalent readings. If using the raw absorbance from a NanoDrop, set path length to 0.1 cm (1 mm) or use the auto-corrected value with 1.0 cm. Short path lengths allow measurement of concentrated samples without dilution.
Common causes: (1) Air bubbles in the cuvette, (2) Particulates or aggregated protein scattering light, (3) Temperature fluctuations affecting buffer absorbance, (4) Dirty cuvette surfaces, (5) Instrument warm-up issues. Allow 15 minutes warm-up, blank with matching buffer, centrifuge samples before measurement, and use clean cuvettes.
Many common buffer components absorb at 280 nm: DTT (significant above 250 nm), imidazole (strong absorbance at 280 nm, problematic for His-tag purification eluates), nucleotides (ATP, GTP), Triton X-100 and NP-40 (aromatic rings). Always blank against the same buffer. Dialyze or desalt protein samples if buffer interference is suspected.
The Warburg-Christian method corrects for nucleic acid contamination using two wavelengths: Protein (mg/mL) = 1.55 × A₂₈₀ - 0.76 × A₂₆₀. This works because nucleic acids absorb more strongly at 260 nm than at 280 nm, while proteins show the opposite pattern. The correction is approximate and works best when contamination is less than 20%.
Yes: µM = (mg/mL × 1000) / MW(Da), or mg/mL = µM × MW(Da) / 1000. For example, 1 mg/mL of a 50 kDa protein = 1000/50000 × 1000 = 20 µM. This conversion requires knowing the molecular weight accurately. Many researchers use mg/mL for protein stocks and µM for binding experiments.
For standard 1-cm cuvettes, the practical detection limit is ~20-50 µg/mL (A₂₈₀ = 0.01-0.05 for typical proteins). NanoDrop instruments can measure 0.1-400 mg/mL with 1-2 µL. For sub-µg/mL concentrations, use fluorescence-based assays (Qubit, NanoOrange) or silver-stained gels for detection.
It depends on the application. Use molar concentration (µM, nM) for binding assays (Kd determination), enzyme kinetics (Km, kcat), and stoichiometric calculations. Use mass concentration (mg/mL) for formulation, column loading calculations, and when MW is uncertain (e.g., glycoproteins with heterogeneous modifications).
Roboculator Team
The Roboculator Team explains calculations, planning tools, and practical formulas in clear language for real-life situations.
How helpful was this calculator?
Be the first to rate!
DNA Copy Number Calculator
Biochemistry Calculators
Protein Molecular Weight Calculator
Biochemistry Calculators
Isoelectric Point Calculator
Biochemistry Calculators
Molar Extinction Coefficient Calculator
Biochemistry Calculators
Peptide Mass Calculator
Biochemistry Calculators
DNA/RNA Molarity Calculator
Biochemistry Calculators